怎样给NI 5112连接一个数字触发信号。



硬件: Modular Instruments>>High-Speed Digitizers (Scopes)

问题:

I have a Digital Trigger Signal that I want to connect to my NI 5112. Should I use the Connector or the PFI1 SMB Jack Connector on the front of my device?

我有一个数字触发信号,我想连接到我的NI 5112上面。我需要用TRIG BNC 接线端还是用设备上的PFI1 SMB接头?



解答: 板卡上面的PFI1,  SMB 接线端 ,是为多种功能的时钟或时钟信号设计的,所以它是你应该连触发信号的部分。数字触发信号应该是 TTL 信号并且至少 10 ns脉宽. TRIG, 板卡下端的母的 BNC接线头 , 是为了连接外部模拟触发信号。TRIG 接线端的信号不能被数字化,因此你不能用它来接你的数字触发信号。

PFI1 接线端也可以产生探头补偿信号。如果你使用PFI1来达到这个目的, 你可以使用 AUX I/O 9-针圆 DIN 接线头,它有一个额外的数字触发线, 在PFI2的第六针。