ポート0以降のポートにてデジタル入出力を試みたときに、エラー200077が発生するのはなぜでしょう。



ハードウェア: Multifunction DAQ (MIO)>>M Series>>PXI-6284, Multifunction DAQ (MIO)>>M Series>>PCI-6220, Multifunction DAQ (MIO)>>M Series>>PCI-6221, Multifunction DAQ (MIO)>>M Series>>PCI-6224, Multifunction DAQ (MIO)>>M Series>>PCI-6229, Multifunction DAQ (MIO)>>M Series>>PCI-6250, Multifunction DAQ (MIO)>>M Series>>PCI-6251, Multifunction DAQ (MIO)>>M Series>>PCI-6254, Multifunction DAQ (MIO)>>M Series>>PXI-6221, Multifunction DAQ (MIO)>>M Series>>PCI-6259, Multifunction DAQ (MIO)>>M Series>>PCI-6280, Multifunction DAQ (MIO)>>M Series>>PCI-6281, Multifunction DAQ (MIO)>>M Series>>PCI-6284, Multifunction DAQ (MIO)>>M Series>>PXI-6220, Multifunction DAQ (MIO)>>M Series>>PXI-6224, Multifunction DAQ (MIO)>>M Series>>PXI-6229, Multifunction DAQ (MIO)>>M Series>>PXI-6250, Multifunction DAQ (MIO)>>M Series>>PXI-6251, Multifunction DAQ (MIO)>>M Series>>PXI-6254, Multifunction DAQ (MIO)>>M Series>>PXI-6259, Multifunction DAQ (MIO)>>M Series>>PXI-6280, Multifunction DAQ (MIO)>>M Series>>PXI-6281, Multifunction DAQ (MIO)>>M Series>>PXI-6289, Multifunction DAQ (MIO)>>M Series>>PCI-6201, Multifunction DAQ (MIO)>>M Series>>PXI-6201, Multifunction DAQ (MIO)>>M Series>>PCI-6289, Multifunction DAQ (MIO)>>M Series>>PXI-6225, Multifunction DAQ (MIO)>>M Series>>PCI-6225, Multifunction DAQ (MIO)>>M Series>>PCIe-6251, Multifunction DAQ (MIO)>>M Series>>PCIe-6259, Multifunction DAQ (MIO)>>M Series>>PCI-6221 (37-Pin), Multifunction DAQ (MIO)>>M Series>>PCI-6230

問題:
MシリーズMIO DAQボードはハードウェアタイミングにおけるデジタル入出力に対応しています。
デバイスのどちらのポートが対応しているのでしょう。
ポート0以外のポートを使用するとエラー200077が発生するのはなぜでしょう。

解決策:
全てのMシリーズデバイスではポート0のみハードウェアタイミングにおけるデジタル入出力に対応しています。
ポート0以外のポートにてハードウェアタイミングにおけるデジタル入出力を試みると、NI-DAQmx Error-200077が発生します。

現在のMシリーズデバイスではハードウェアタイミングでのデジタル入出力に8または32ライン対応しています。
デバイスの詳細について以下リンク先のMシリーズヘルプ(ユーザマニュアルとSpecifications)をご参照下さい。


関連リンク:
M Series Help (User Manual and Specifications)

添付:





報告日時: 09/14/2005
最終更新日: 01/25/2008
ドキュメントID: 3PD89BRM