如果在测试板卡时仅仅使用到了Verify ,那么校准日期将会保持不变。仅在使用Verify and Adjust时,校准日期才会发生改变。除非执行了调整的操作,否则EEPROM不会被重新写入并且校准日期也会保持不变。以下是能够在 Calibration Executive Manual 中找到的内容(链接如下):
Verify :将测量所得到的设备性能和已知测量不确定度的外化标准相比较,以确认设备满足或超过其规格
Adjust:通过调整校准常数,修正设备的测量误差,并将新的校准常数存储在EEPROM中。主机会读取校准常数,软件利用并修改数据上的错误,同时将校准时间呈现给用户。
执行Verify and Adjust 将会保证最新的校准时间得到更新。然而,利用LabVIEW或者ANSI C的编程方法同样可能在没有调整的前提下,改变这个校准日期。在LabVIEW中使用DAQmx Initialize External Calibration.vi and the DAQmx Close External Calibration.vi完成这个操作。查看如下所示的vi片段。它将会打卡并关闭设备的校准参考,在没有进行任何调整的情况下更新校准日期。
注意:以上的方法不会更新NI 9205或者NI 9206的校准日期。
|