From Friday, April 19th (11:00 PM CDT) through Saturday, April 20th (2:00 PM CDT), 2024, ni.com will undergo system upgrades that may result in temporary service interruption.

We appreciate your patience as we improve our online experience.



當我跑 NI435x Getting Started 範例時我會得到一個關於Accessory的錯誤訊息



硬體: Modular Instruments>>High-Precision DAQ (Dataloggers)

問題:

當我跑NI435x Getting Started.vi 時會出現 "Error -1306 that is explained as NI435X Check Accessory."
的錯誤訊息.

或者

當在Visual Basic跑NI435x Getting_Started 範例時會出現 "Error -1316 in NI435X_Set_Auto_Zero: ERROR: No autozero on accessory." 的錯誤訊息.



解答:

連接在NI435x的Accessory需要在MAX底下被設定才能支援這些功能. 如要設定Accessory請依以下的步驟:

1. 開啟MAX並按F5做更新.

2. 展開 Devices and Interfaces的欄位.

3. 按右鍵點選435x並選"Properties".

4. 選擇Accessory欄位.

5. 在下拉式選單內選擇您的Accessory

此時您的Accessory與量測應已被正確的設定.



相關連結:

附加檔案:





報告日期: 04/03/2003
最後更新: 12/18/2008
文件偏號: 2W2DONZO