niTclk同步VI产生错误-250033



主要软件: NI Real-Time Hypervisor
主要软件版本: 2.0
主要软件修正版本: N/A
次要软件: N/A
硬件: Modular Instruments

问题: 我尝试使用Tclk技术同步多块PXI板卡,但是得到以下错误。这是为什么?
Error -250033 occurred at niTClk Synchronize.vi

Possible reason(s):

NI-TClk committed a session, and the instrument driver reported an error.

No registered trigger lines could befound between the devices in the route.


解答: 这个错误显示的是基于Windows的NI实时管理程序的环境。当基于Windows的NI实时管理程序管理电脑的硬件资源时与NI Tclk同步驱动的路由功能发生冲突。

可以使用T-clk采用以下方法:
shared sample clock and start trigger
或者 booting into the Windows environment without using the Hypervisor option.

相关链接: Knowledge Base 4Y8CEGL3: Basic NI Real-Time Hypervisor Trigger Routing
Knowledge Base 4ZUC5K0S: Advanced PXI Trigger in NI Real-Time Hypervisor
Knowledge Base 4YFEI84I: Frequently Asked Questions (FAQ) for the NI Real-Time Hypervisor
Knowledge Base 4H3ERMG3: Error Code 250033: NI-TClk Committed a Session, and the Instrument Driver Reported an Error.
NI Developer Zone Tutorial: NI Real-Time Hypervisor for Windows Walkthrough
NI Developer Zone Tutorial: Synchronization Explained