Archived: WLS/ENET-9163 完了まで待機でエラーコード-200560が発生します
This document has been archived and is no longer updated by National Instruments
ハードウェア: PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers, PXI/CompactPCI>>Controllers
問題: WLS/ENET-9163をDAQmxタスクを開始とDAQmx読み取りの間にDAQmx基準トリガを配置し、DAQmx完了まで待機を用いてDAQmxタスクを停止の前に計測をきちんと終了させたいとするとタイムアウトが発生し、エラー-200560が発生します。

このエラーはトリガがとれ、すべてのデータがデバイスから集録されていても発生します。
解決策: このエラーを回避するためにはDAQmx読み取りのプロパティノードのチャンネルごとの有効なサンプル数(Available Samples Per Channel)をDAQmx完了まで待機の前に置くことで回避できます。もしプロパティノードの前に集録が終わっている場合、エラーを返さずにDAQmx完了まで待機VIが完了します。
修正された添付VIをご参照下さい。
関連リンク: KnowledgeBase 4QNAFMD2: Cannot Connect to NI WLS/ENET-9163 Devices on W2K with IE5
KnowledgeBase 4QNB43D2: WLS/ENET-9163 is Undiscoverable Over Ethernet with Static IP
添付:
AI_-_Ext_Ref_Trigger.vi
報告日時: 11/05/2008
最終更新日: 03/18/2009
ドキュメントID: 4R4AEQD2