當我嘗試執行self-test或是使用安裝在LabVIEW Real-time下的DAQ裝置與NI-DAQmx 8.7.1時,我得到下面的錯誤訊息:
Error -88716 occurred at Test Panel
Possible Reason(s):
The requested operation failed because the versions of a driver installed on the local and the remote system are different.
Update the driver so the versions on the two systems are the same.
NI-DAQmx 8.7.1f2 有已知的問題會導致這個錯誤.NI-DAQmx 8.7.1f3 解決了這個問題.如果你得到這個錯誤並且確認你是安裝NI-DAQmx 8.7.1f2,那麼請下載並安裝NI-DAQmx 8.7.1f3.當下載並安裝NI-DAQmx 8.7.1f3將會更新 LabVIEW Real-time target上面的NI-DAQmx.
要看目前安裝在主機端的NI-DAQmx版本,請開啟Measurement and Automation, 展開My system底下的Software, 選擇NI-DAQmx來判斷現在的版本.
|