From 11:00 PM CDT Friday, Nov 8 - 2:30 PM CDT Saturday, Nov 9, ni.com will undergo system upgrades that may result in temporary service interruption.

We appreciate your patience as we improve our online experience.



NI-DAQ 7.0或7.1無法正常驅動AT介面E系列元件



主要軟體:
主要軟體版本: 7.0
主要軟體修正版本: 7.2
次要軟體: N/A
硬體: Multifunction DAQ (MIO)>>E Series, Legacy DAQ Devices>>Multifunction DAQ>>AT-MIO-16XE-10, Legacy DAQ Devices>>Multifunction DAQ>>AT-MIO-16E-1, Legacy DAQ Devices>>Multifunction DAQ>>AT-MIO-64E-3, Legacy DAQ Devices>>Multifunction DAQ>>AT-MIO-16DE-10, Multifunction DAQ (MIO)>>E Series>>AT-MIO-16E-10, Multifunction DAQ (MIO)>>E Series>>AT-MIO-16XE-50, Multifunction DAQ (MIO)>>E Series>>AT-MIO-16E-2, Legacy DAQ Devices>>Multifunction DAQ>>AT-AI-16XE-10

問題: 當我使用NI-DAQ 7.0或7.1驅動我的AT介面E系列元件時,我在MAX(Measurement & Automation Explorer)的測試面板得到以下的錯誤訊息:


Device is not responding to the first DMA channel


但是裝置管理員顯示三個DMA(直接記憶體存取)的通道皆正常,但是MAX卻顯示錯誤訊息,並且第一個DMA通道無回應。我還得到以下的錯誤訊息:
Error 10800 (timeout) in DAQ_Op
對於所有AT介面E系列元間接無法用NI-DAQ 7.0或7.1驅動。我要如何才可以解決這個問題呢?

解答: 您還是可以利用AT介面E系列元件擷取資料,但是卻無法使用直接記憶體存取通道。而PCI和PXI介面的E系列元件卻不會有任何影響而正常驅動。請升級NI-DAQ 7.2(參可以下連結)來解決此問題。

相關連結:
NI-DAQ Version 7.2 for Windows 2000/NT/XP


附加檔案:





報告日期: 05/14/2004
最後更新: 03/10/2005
文件偏號: 34F82RTY