Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
This Document is not yet Rated  Rate this Document

Error -200077 When Using an NI 9263 with Calibration Executive 3.2

Primary Software: Calibration Executive
Primary Software Version: 3.1
Primary Software Fixed Version: N/A
Secondary Software: N/A
Hardware: CompactRIO>>Analog Output Modules>>NI 9263, CompactRIO>>Analog Output Modules>>NI 9263

Problem: When I attempt to calibrate an NI 9263 (cRIO/cDAQ module) with Calibration Executive 3.2, I'm presented with error -200077.

Solution: The extended information for this error is:

Requested value is not a supported value for this property.
Property: AO.TermCfg
You Have Requested: Differential
You Can Select: RSE


The procedure written for Calibration Executive 3.2 attempts to set the terminal configuration to an impossible value.  For further assistance with this problem, please contact NI Technical Support.

Related Links:

Attachments:





Report Date: 04/09/2007
Last Updated: 04/16/2007
Document ID: 488HJ1D3

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit