Error -200077 When Using an NI 9263 with Calibration Executive 3.2 Primary Software: Calibration ExecutivePrimary Software Version: 3.1 Primary Software Fixed Version: N/A Secondary Software: N/A Hardware: CompactRIO>>Analog Output Modules>>NI 9263, CompactRIO>>Analog Output Modules>>NI 9263
Problem: When I attempt to calibrate an NI 9263 (cRIO/cDAQ module) with Calibration Executive 3.2, I'm presented with error -200077. Solution: The extended information for this error is: Requested value is not a supported value for this property. Property: AO.TermCfg You Have Requested: Differential You Can Select: RSE The procedure written for Calibration Executive 3.2 attempts to set the terminal configuration to an impossible value. For further assistance with this problem, please contact NI Technical Support. Related Links: Attachments:
Report Date: 04/09/2007 Last Updated: 04/16/2007 Document ID: 488HJ1D3 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||