Error -1074118518 or -1074111850 When Streaming Waveforms Using NI-FGEN

Hardware: Signal Sources (AWG/FG/AO)

Problem:
I receive Error -1074118518 or Error -1074118520 when I try to stream waveform data using NI-FGEN. What is causing these errors?

Solution:
Error -1074118518

The NI-FGEN driver has a minimum required waveform size. For device specific minimum waveform sizing see the detailed specifications of your device. These can be found from the product page under the "Resources" tab. If you have selected a value that does not meet this minimum requirement, you will need to change the Waveform Size value.

Error -1074118520

The NI-FGEN driver requires that the quotient of the Waveform Size divided by the Data Transfer Block Size be an integer value. That is, there cannot be a fractional number of data blocks in the waveform. If you have selected waveform size that is not an integer multiple of the block size you must change either the Waveform Size or Data Transfer Block Size value.