Multiple Digital Triggers Using Real-Time System Integration (RTSI)

Primary Software: Driver Software>>NI-DAQ
Primary Software Version: 7.3
Primary Software Fixed Version: N/A
Secondary Software: N/A
Hardware: Digital I/O (DIO)>>High-Speed>>AT-DIO-32HS

Problem:
Multi-function I/O (MIO) boards have at most one hardware digital trigger. How can I use National Instruments products to trigger a data acquisition system with more than one digital trigger?

Solution:
You can use a high-speed digital I/O board (such as the DIO-32HS) to accomplish this. Connect the RTSI bus between the DIO-32HS (6533 Family) and the MIO board. Configure the DIO-32HS for change detection mode and route the REQ signal to the RTSI. Configure the MIO board's digital trigger source to use the REQ signal, which is now available on the RTSI (from the DIO-32HS). Finally, wire your digital trigger sources to the input lines on the DIO-32HS board. The example attached below is written in LabVIEW. It configures the first port for change detection mode, which asserts the REQ line when any digital input line transitions between logical states.

Related Links:
Developer Zone Example: Multiple Digital Triggers Using RTSI between MIO Board and DIO Board
Developer Zone Example: Acquire N Scans with Multiple Digital Triggers (Efficiently)
Developer Zone Example: Performing Single-Shot Waveform Acquisition with Multiple Analog Software Triggers

Attachments:


multiple digital triggers with 32HS.zip


Report Date: 04/03/2000
Last Updated: 09/16/2007
Document ID: 1W2K474U

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit