LFH200 Bare Wire Cables with Date Codes 3707 and 5007 Have Incorrect Pinouts Hardware: Switches>>Multiplexer/Matrix>>PXI-2575, SCXI>>Modules>>SCXI-1175, Switches>>Multiplexer/Matrix>>PXIe-2575, Switches>>General-Purpose>>PXI-2569, Switches>>General-Purpose>>PXI-2570, SCXI>>Modules>>SCXI-1169
Problem: On the LFH200 Bare Wire cable, Leg J2, Leg J3, Leg J6 and Leg J7 have mismatched wire color codes when compared with the pin assignment table given in the LFH200 Cable Installation Instructions. This difference causes incorrect signal routing. Why is my LFH200 Bare Wire cable not matching the pin assignment table in the installation instructions? Solution: Beginning at the end of 2007 and continuing until January 2008, there were units of the LFH200 cable where several legs of the cable had wires connected in reverse of what they should be. Because of this, the table in the LFH200 Installation Instructions does not completely match up with the wires on the LFH200 cable. Affected cables have date codes of 3707 and 5007. The date code can be found on the fourth line of the cable label. See below for a modified pinout for the mis-wired LFH200 cables. National Instruments is committed to your total satisfaction. See the link below to contact us. Please accept our apology for the inconvenience. Related Links: Products and Services: NI LFH200 Cables Attachments:
Report Date: 02/06/2008 Last Updated: 03/13/2008 Document ID: 4I5G6BLE |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||