Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
This Document is not yet Rated  Rate this Document

Error-17338 When Closing Calibration Executive 3.3

Primary Software: Calibration Executive
Primary Software Version: 3.3
Primary Software Fixed Version: N/A
Secondary Software:

Problem:
Error -17338 occurs when closing Calibration Executive after a Calibration procedure when LV 8.6 RTE is also installed with Calibration Executive 3.3. How can I get rid of this error?



Solution:
This is the known issue in Calibration Executive 3.3. The workaround to eliminate this error is by modifying  the following line in the TestExec.ini file.  TestExec.ini can be found in
\Program Files\National Instruments\TestStand 4.0\Cfg\TestExec.ini

Change:
_FlexGAdp_RTSPath = C:\\Program Files\\National Instruments\\Shared\\LabVIEW Run-Time\\8.x\\lvrt.dll

To:
_FlexGAdp_RTSPath = "C:\\Program Files\\National Instruments\\Shared\\LabVIEW Run-Time\\8.5\\lvrt.dll





Related Links:
NI Calibration Services

Attachments:





Report Date: 06/05/2009
Last Updated: 06/05/2009
Document ID: 4Y4ACABH

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit