Error-17338 When Closing Calibration Executive 3.3Primary Software: Calibration ExecutivePrimary Software Version: 3.3 Primary Software Fixed Version: N/A Secondary Software:
Problem: Error -17338 occurs when closing Calibration Executive after a Calibration procedure when LV 8.6 RTE is also installed with Calibration Executive 3.3. How can I get rid of this error? ![]() Solution: This is the known issue in Calibration Executive 3.3. The workaround to eliminate this error is by modifying the following line in the TestExec.ini file. TestExec.ini can be found in \Program Files\National Instruments\TestStand 4.0\Cfg\TestExec.ini Change: _FlexGAdp_RTSPath = C:\\Program Files\\National Instruments\\Shared\\LabVIEW Run-Time\\8.x\\lvrt.dll To: _FlexGAdp_RTSPath = "C:\\Program Files\\National Instruments\\Shared\\LabVIEW Run-Time\\8.5\\lvrt.dll Related Links: NI Calibration Services Attachments:
Report Date: 06/05/2009 Last Updated: 06/05/2009 Document ID: 4Y4ACABH |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||