M系列板卡管脚与SCXI-1349和SCXI-1346旁路接口管脚的对应关系。
硬件:    SCXI>>Terminal Blocks
问题:    M系列板卡的68针管脚与SCXI-1349和SCXI-1346的50针旁路接口管脚是怎样对应的?
解答:     | 68-pin M-Series Line Name |  68-pin M-Series Pin Number |  50-pin Breakout Connector Pin Number | 
 | AIGND |  24* |  1 | 
 | AIGND |  24* |  2 | 
 | AI 0 |  68 |  3 | 
 | AI 8 |  34 |  4 | 
 | AI 1 |  33 |  5 | 
 | AI 9 |  66 |  6 | 
 | AI 2 |  65 |  7 | 
 | AI 10 |  31 |  8 | 
 | AI 3 |  30 |  9 | 
 | AI 11 |  63 |  10 | 
 | AI 4 |  28 |  11 | 
 | AI 12 |  61 |  12 | 
 | AI 5 |  60 |  13 | 
 | AI 13 |  26 |  14 | 
 | AI 6 |  25 |  15 | 
 | AI 14 |  58 |  16 | 
 | AI 7 |  57 |  17 | 
 | AI 15 |  23 |  18 | 
 | AISENSE |  62 |  19 | 
 | AO 0 |  22 |  20 | 
 | AO 1 |  21 |  21 | 
 | APFI 0 |  20 |  22 | 
 | AOGND |  54,55 |  23 | 
 | DGND/PFI 15/P2.7 |  4** |  24 | 
 | P0.0 |  52*** |  25 | 
 | P0.4 |  19*** |  26 | 
 | P0.1 |  17*** |  27 | 
 | P0.5 |  51 |  28 | 
 | P0.2 |  49*** |  29 | 
 | P0.6 |  16 |  30 | 
 | P0.3 |  47 |  31 | 
 | P0.7 |  48 |  32 | 
 | DGND |  4** |  33 | 
 | +5V |  8,14 |  34 | 
 | +5V |  8,14 |  35 | 
 | PFI 11/P2.3 |  46 |  36 | 
 | PFI 10/P2.2 |  45 |  37 | 
 | PFI 0/P1.0 |  11 |  38 | 
 | PFI 1/P1.1 |  10 |  39 | 
 | PFI 2/P1.2 |  43 |  40 | 
 | PFI 3/P1.3 |  42 |  41 | 
 | PFI 4/P1.4 |  41 |  42 | 
 | PFI 13/P2.5 |  40 |  43 | 
 | PFI 5/P1.5 |  6 |  44 | 
 | PFI 6/P1.6 |  5 |  45 | 
 | PFI 7/P1.7 |  38 |  46 | 
 | PFI 8/P2.0 |  37 |  47 | 
 | PFI 9/P2.1 |  3 |  48 | 
 | PFI 12/P2.4 |  2 |  49 | 
 | PFI 14/P2.6 |  1 |  50 | 
*同时和M系列板卡的27, 29, 32, 56, 59, 64, 以及 67针相连。
**同时和M系列板卡的7, 9, 12, 13, 15, 18, 35, 36, 39, 44, 50, 以及 53 针相连。
***被SCXI机箱占用的数字资源。
注意:SCXI-1349和SCXI-1346旁路接口应配合SCXI-1180模块使用。
相关链接:   National Instruments: SCXI Support Page
Product Manuals: SCXI-1349 Shielded Cable Installation Guide
Product Manuals: SCXI-1346 Shielded Multi-Chassis Cable Adapter Installation Guide
Product Manuals: SCXI-1180/1181 User Manual
附件:   
报告日期:   11/08/2004
最近更新:   11/18/2004
文档编号:   3F7DNTNY
 
    
        
            
                
                
                    Other Support Options
                    Ask the NI Community
                    Collaborate with other users in our discussion forums
                    
                
                
                Request Support from an Engineer
                A valid service agreement may be required, and support options vary by country.