Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
This Document is not yet Rated  Rate this Document

Why Do I Get Error -200452 When I Run Write Dig Chan-Int Clk-Dig Start.vi Using a USB-6259

Primary Software: LabVIEW Development Systems>>LabVIEW Base Package
Primary Software Version: 8.6.1
Primary Software Fixed Version: N/A
Secondary Software: N/A
Hardware: Multifunction DAQ (MIO)>>M Series>>USB-6259 BNC

Problem:
I receive Error -200452 with a possible reason of Specified property node is not supported by the device or is not applicable to the test. Is there an example vi similar to Write Dig Chan-Int Clk-Dig Start that would work with my NI 6259 without giving me this error?



Solution:
In the LabVIEW Getting Started Window, select Find Examples... Double-Click Hardware Input and Output»DAQmx»Digital Generation»Write Dig Chan - Ext Clk.vi







The reason why you are receiving this error is because the digital subsystem for the M Series devices does not have its own dedicated internal timing engine. Therefore, a sample clock must be provided from another subsystem on the device or an external source.

Related Links:
Product Manuals: NI 625x Specifications
Knowledgebase 4PSF9EE9: Where Can I Find LabVIEW or LabWindows/CVI Example Programs Specifically for my Hardware?

Attachments:





Report Date: 04/15/2009
Last Updated: 10/22/2009
Document ID: 4WEHICP4

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit