NI 4451 Fails When Using the Acquire N Scans Example for Visual BasicPrimary Software: Measurement Studio>>Visual Basic SupportPrimary Software Version: 7.0 Primary Software Fixed Version: N/A Secondary Software: N/A Hardware: Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4451, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4452, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PCI-4551, Modular Instruments>>Dynamic Signal Acquisition and Analysis (DSA)>>PXI-4552
Problem: The Acquire N Scans example in Visual Basic reports an error on OneShot acquisition using an NI 4451 or NI 4452. The error is -10403 (deviceSupportError) and occurs when starting a OneShot acquisition. After proceeding through the error dialog box, the data is actually acquired fine. Solution: If you select continuous mode instead of OneShot, the error does not occur. This problem has been resolved in the ComponentWorks 2.0. You can disable ExceptionOnError for the AI control to avoid this error message. Related Links: Product Manual: PCI-4451/4452/4453/4454 User Manual (March 2000) KnowledgeBase 1ME7AU0O: Problems with Waveform Generation on Dynamic Signal Analyzer (DSA) Boards in C through NI-DAQ KnowledgeBase 1F9CDEDO: My Dynamic Signal Analyzer (DSA) Instrument Fails in Example Cont Acq&Chart (hw timed).vi Attachments:
Report Date: 03/30/1998 Last Updated: 04/22/2007 Document ID: 17TEI7JV |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||