Why does the Auto-Detect Fail to Recognize my SCXI-1163R?Hardware: SCXI>>Modules>>SCXI-1163R
Problem: I am controlling a SCXI chassis with an M-Series DAQ device. Currently I have the chassis properly configured in MAX with the appropriate communicating device selected and it auto-detects and recognizes all modules except my SCXI 1163R. I have ensured that all the jumpers are in their default configurations, yet the device is still not detected. Solution: The SCXI-1163R’s default jumper settings configure the module to be controlled with a Digital IO device. The M Series device that is being used to control the SCXI module is an MIO device, so you need to change three of the jumpers. Change your jumper configuration to the following:
Related Links: Product Manuals: NI SCXI-1163R Jumpers Note to Users Attachments:
Report Date: 04/22/2008 Last Updated: 04/24/2008 Document ID: 4KLC2V9K |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||