Traditional NI-DAQ (Legacy) Document Includes Incorrect Units for SCXI-1124 Current Calibration Primary Software: Driver Software>>NI-DAQPrimary Software Version: 7.0 Primary Software Fixed Version: N/A Secondary Software: N/A Hardware: SCXI>>Modules>>SCXI-1124
Problem: The Traditional NI-DAQ (Legacy) Function Reference Manual indicates that current measurements for SCXI-1124 current calibration should be entered in amps. However, shouldn't current values be expressed in milliamps? Solution: The SCXI-Cal_Constants documentation in the Traditional NI-DAQ (Legacy) Function Reference Manual does include incorrect information. When performing a two-point calibration of an SCXI-1124, values for current should be expressed in milliamps (mA), not amps.Related Links: Attachments:
Report Date: 09/12/2006 Last Updated: 11/01/2007 Document ID: 41B9RM1V |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||