Traditional NI-DAQ (Legacy) Document Includes Incorrect Units for SCXI-1124 Current Calibration

Primary Software: Driver Software>>NI-DAQ
Primary Software Version: 7.0
Primary Software Fixed Version: N/A
Secondary Software: N/A
Hardware: SCXI>>Modules>>SCXI-1124

Problem:
The Traditional NI-DAQ (Legacy) Function Reference Manual indicates that current measurements for SCXI-1124 current calibration should be entered in amps. However, shouldn't current values be expressed in milliamps?

Solution:
The SCXI-Cal_Constants documentation in the Traditional NI-DAQ (Legacy) Function Reference Manual does include incorrect information. When performing a two-point calibration of an SCXI-1124, values for current should be expressed in milliamps (mA), not amps.