Why is My SCXI-1160/1161 Switch Modules Not Detected in MAX?Hardware: Switches>>General-Purpose>>SCXI-1160
Problem: I have cabled my SCXI-1160 to a PCI-4021 switch controller via the 1362 cable assembly kit. However, Measurement and Automation Explorer (MAX) does not detect my SCXI-1160 or any of the other devices in the chassis. I have the latest version of the NI-DAQ and NI-Switch drivers installed. Why is my SCXI-1160 switch module not detected in MAX? Solution: The most common cause of this behavior is incorrect jumper settings. Jumpers W2, W3, and W4 on the SCXI-1160/1161 have two labeled positions -- MIO and DIO. In order for the SCXI-1160/1161 to communicate with the PCI-4021, these three jumpers must all be set to the MIO position. If they are set to DIO, the SCXI-1160/1161 will not be detected in MAX. Also verify that jumpers W1 and W5 are set in the correct positions according to your chassis setup: For a single chassis, set W1 and W5 to 1-2. For multiple chassis:
For futher information on the SCXI-1160/1161, please see the NI Switches Help by clicking on the Manuals link below, entering 1160 as the keyword, and clicking on the Go button. Related Links: Manuals Attachments:
Report Date: 02/25/2004 Last Updated: 02/26/2004 Document ID: 36ODCNCW |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||