The NI PXI-5600 RF Downconverter Does Not Show up in Measurement and Automation Explorer (MAX) on the Real-Time (RT) Target When Using NI-DAQ 8.7. Primary Software: Driver Software>>NI-RFSAPrimary Software Version: 2.0.7 Primary Software Fixed Version: N/A Secondary Software: N/A
Problem: If a customer installs NI-DAQ 8.7 and then NI-RFSA 2.0.7, RT support for the NI PXI-5600 won't work (i.e. the NI PXI-5600 will not be recognized as a device) without an additional step. Solution: After downloading NI-RFSA 2.0.7 to their RT target, the customer will need to redownload NI-DAQ to the RT target. After rebooting the RT system, the NI 5600 should appear in MAX. Related Links: Attachments:
Report Date: 03/04/2008 Last Updated: 03/11/2008 Document ID: 4J3GQHY5 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
