Application Developed with an E-Series DAQ Device gives Error -10007 when running with DAQCard-700 Primary Software: LabVIEW Development Systems>>Professional Development SystemPrimary Software Version: 7.1 Primary Software Fixed Version: N/A Secondary Software: N/A Hardware: Legacy DAQ Devices>>Multifunction DAQ>>DAQCard-700
Problem: Why do I get Error -10007 when running an application with my DAQCard-700? The same program works fine with an E-Series DAQ device. Solution: When performing multichannel data acquisition, the scan counter on the DAQCard-700 decrements from the highest numbered channel (selected by the user) through channel 0. The E-Series board however, can accept a scan list reading the analog channels in any order. Changing the scan list to the appropriate order when using the DAQCard-700 should avoid the error. Related Links: Product Manual:DAQCard-700 User Manual KnowledgeBase 1DFCE80O: Why Can't I Define an Arbitrary Scan List with Lab or 1200-Series Hardware? Attachments:
Report Date: 01/03/2005 Last Updated: 06/04/2008 Document ID: 3H2CQ8NK |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||