Error -200077 (0xFFFCF273) During Thermocouple Measurements with SCB-68 or SCXI
Primary Software: Utility Software>>Measurement & Automation Explorer
Primary Software Version: 3.1
Primary Software Fixed Version: N/A
Secondary Software: N/A
Hardware: Counter/Timers (TIO)>>Connectivity>>SCB-68, SCXI>>Modules>>SCXI-1102
Problem: Why is using a DAQ device with the SCB-68 or SCXI-1102 for Thermocouple Measurements with Built-In CJC Compensation Yield -200077 Error?
Solution: SCXI-1102- Launch Measurement and Automation Explorer and navigate to the SCXI-1102 module
- Right-click the SCXI-1102, and select Properties
- Select the Accessory Tab
- Ensure that the Accessory in the drop down menu has CJC capabilities (such as the SCXI-1303)
SCB-68 - Launch Measurement and Automation Explorer and navigate to the DAQ device that is attached to the SCB 68
- Right-click on the DAQ device and select Properties
- Select the Accessory Tab
- Ensure that under Accessory drop down box the SCB 68 is selected, then left-click the Configure button
- Under the CJC Configuration tab, ensure that the " Temperature Reference Enabled " checkbox is checked.
- Ensure that temperature reference sensor is enabled on the SCB 68. This can be found on page 4-3 of SCB 68 User Manual (Linked below)
Related Links: SCB 68 User Manual
KnowledgeBase 3UKG06X1: Conflicting Range Values in Thermocouple Task Cause Error -200077