How can I use my SCC-A10 Modules with VirtualBench-Scope?

Primary Software: Other NI Software>>VirtualBench
Primary Software Version: 2.51
Primary Software Fixed Version: N/A
Secondary Software: N/A

Problem: If you configure the SCC-2345 as an accessory for your Data Acquisition board and a SCC-A10 for a particular channel in the Measurement and Automation Explorer, then the NI-DAQ driver will automatically return a voltage reading on the order of +/- 100 V. This causes VirtualBench to report an over range error.

Solution: In order to avoid this overrange problem, you can still have the SCC-2345 configured as an accessory in the Measurement and Automation Explorer (or the NI-DAQ Configuration Utility for earlier versions of NI-DAQ), however do not specify a particular channel as having the SCC-A10 installed. Then use the Probe Attenuation setting in the VirtualBench-Scope Channel Settings and specify a probe attenuation of 10. Your signal will now be correct.

Related Links:

Attachments:





Report Date: 08/25/1999
Last Updated: 12/08/2004
Document ID: 1OOB56TS

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit