Changing the Resman Delay of the GPIB-VXI/C Hardware: GPIB>>Plug-in Controllers>>GPIB-VXI/C
Problem: How do I change the Resman delay of the GPIB-VXI/C? Solution: The GPIB-VXI/C is factory configured such that Resman will run immediately following bootup (usually about 2 seconds). When a VXI chassis is powered up, the instruments in the chassis perform a self test which sometimes fails to complete before Resman runs. Therefore, Resman may not see all the instruments in the chassis, and you may need to increase the Resman delay to allow time for the instruments to complete their self tests. The instructions below will show you how to change the Resman delay. To change the Resman delay, you must modify the non-volatile configuration through the serial interface (refer to the related link below for information on how to set up Hyperterminal for the GPIB-VXI/C). With the serial cable connected and Hyperterminal running, power on the chassis. You will see the output of Resman followed by a prompt.
GPIB-VXI Nonvolatile Configuration Main Menu ================================================ 1). Read In Nonvolatile Configuration 2). Print Configuration Information 3). Change Configuration Information 4). Set Configuration to Factory Settings 5). Write Back (Save) Changes 6). Quit Configuration Choice (1-6): 3
GPIB-VXI Nonvolatile Configuration Changer (C) 1995 National Instruments ================================================== 0). Edit Local Register Configuration 1). Edit pSOS Configuration 2). Edit VXI Interrupt Handler Logical Addresses 3). Edit Resource Manager Configuration 4). Edit Servant Area and DC Configuration 5). Edit FAILED Device Handling Mode 6). Edit GPIB Configuration 7). Edit Default CI Configuration 8). Edit Resident CI Base Locations 9). Edit CI User Configuration Variables Q). Quit Editor Choice (0-9,Q): 3
Seconds to wait before starting Resource Manager (default 0x0005): 3 A24 Base Address to Start Assigning Memory Map (default 0x200000): A32 Base Address to Start Assigning Memory Map (default 0x20000000):
GPIB-VXI Nonvolatile Configuration Changer (C) 1995 National Instruments ================================================== 0). Edit Local Register Configuration 1). Edit pSOS Configuration 2). Edit VXI Interrupt Handler Logical Addresses 3). Edit Resource Manager Configuration 4). Edit Servant Area and DC Configuration 5). Edit FAILED Device Handling Mode 6). Edit GPIB Configuration 7). Edit Default CI Configuration 8). Edit Resident CI Base Locations 9). Edit CI User Configuration Variables Q). Quit Editor
GPIB-VXI Nonvolatile Configuration Main Menu (C) 1995 National Instruments ================================================ 1). Read In Nonvolatile Configuration 2). Print Configuration Information 3). Change Configuration Information 4). Set Configuration to Factory Settings 5). Write Back (Save) Changes 6). Quit Configuration
Related Links: KnowledgeBase 3YSDN4DW: Setting Up HyperTerminal to Access the GPIB-VXI/C Attachments: