High Speed Digital Calibration Requirements

Hardware: Digital I/O (DIO)>>High-Speed>>PXI-6542

Problem:
I can see from the National Instruments calibration page that 655x devices can be calibrated, however the 654x and 656x device cannot. When will calibration for these device be available, or do they not require calibration?

Solution:
Both the 654x and 656x devices do not require calibration. Through design, testing and validation over lot variances and temperature the timing circuitry is guaranteed not to drift out of specification.

The same applies to the voltage calibration of the 654x devices. Although voltages levels can be changed to be compatible with different logic families, it should not drift out of specification and does not require calibration.