Why do I Get Error -50150 when Enabling Records > Memory and TIS?

Hardware: Modular Instruments>>High-Speed Digitizers (Scopes)>>PXI-5152, Modular Instruments>>High-Speed Digitizers (Scopes)>>PCI-5154, Modular Instruments>>High-Speed Digitizers (Scopes)>>PXI-5153, Modular Instruments>>High-Speed Digitizers (Scopes)>>PXI-5154, Modular Instruments>>High-Speed Digitizers (Scopes)>>PCI-5152, Modular Instruments>>High-Speed Digitizers (Scopes)>>PCI-5153

Problem:
I am trying to perform a multi-record acquisition using Time-Interleaved Sampling (TIS) with an NI-515x High-Speed Digitizer. I have both Records > Memory and TIS enabled using property nodes, but when I run my VI I get error -1074108567 with status -50150.

Solution:
This is a known issue which has been fixed in NI-SCOPE 3.5.1. In earlier versions of NI-SCOPE this error will occur when fetching more than one record at a time with both Records > Memory and TIS enabled as illustrated below.

.

Upgrading to NI-SCOPE 3.5.1 or later will resolve the error. Visit Drivers and Updates to download the latest NI-SCOPE driver.