為什麼我在port 0以外的數位訊號埠,使用關聯性數位I/O(Correlated Digital I/O),會得到Error -200077的錯誤訊息?



硬體: Multifunction DAQ (MIO)>>M Series>>PXI-6284, Multifunction DAQ (MIO)>>M Series>>PCI-6220, Multifunction DAQ (MIO)>>M Series>>PCI-6221, Multifunction DAQ (MIO)>>M Series>>PCI-6224, Multifunction DAQ (MIO)>>M Series>>PCI-6229, Multifunction DAQ (MIO)>>M Series>>PCI-6250, Multifunction DAQ (MIO)>>M Series>>PCI-6251, Multifunction DAQ (MIO)>>M Series>>PCI-6254, Multifunction DAQ (MIO)>>M Series>>PXI-6221, Multifunction DAQ (MIO)>>M Series>>PCI-6259, Multifunction DAQ (MIO)>>M Series>>PCI-6280, Multifunction DAQ (MIO)>>M Series>>PCI-6281, Multifunction DAQ (MIO)>>M Series>>PCI-6284, Multifunction DAQ (MIO)>>M Series>>PXI-6220, Multifunction DAQ (MIO)>>M Series>>PXI-6224, Multifunction DAQ (MIO)>>M Series>>PXI-6229, Multifunction DAQ (MIO)>>M Series>>PXI-6250, Multifunction DAQ (MIO)>>M Series>>PXI-6251, Multifunction DAQ (MIO)>>M Series>>PXI-6254, Multifunction DAQ (MIO)>>M Series>>PXI-6259, Multifunction DAQ (MIO)>>M Series>>PXI-6280, Multifunction DAQ (MIO)>>M Series>>PXI-6281, Multifunction DAQ (MIO)>>M Series>>PXI-6289, Multifunction DAQ (MIO)>>M Series>>PCI-6201, Multifunction DAQ (MIO)>>M Series>>PXI-6201, Multifunction DAQ (MIO)>>M Series>>PCI-6289, Multifunction DAQ (MIO)>>M Series>>PXI-6225, Multifunction DAQ (MIO)>>M Series>>PCI-6225, Multifunction DAQ (MIO)>>M Series>>PCIe-6251, Multifunction DAQ (MIO)>>M Series>>PCIe-6259, Multifunction DAQ (MIO)>>M Series>>PCI-6221 (37-Pin), Multifunction DAQ (MIO)>>M Series>>PCI-6230

問題:
我知道M系列的DAQ卡片支援關聯性(即硬體時脈)數位I/O功能。請問哪一個數位不支援此功能?為什麼使用非Port 0的數位埠,會得到-200077的錯誤訊息?

解答:
所有M系列的DAQ卡片只有Port 0支援關聯性數位I/O。如果您有在其他的數位埠使用此功能,就會得到NI-DAQmx的錯誤訊息-200077。現在的M系列卡片提供8個或32個通道硬體時脈數位I/O。有關於元件更進一步的訊息,可以參考以下連結M series help(User Manual and Specifications)。

相關連結:
M Series Help (User Manual and Specifications)

附加檔案:





報告日期: 09/14/2005
最後更新: 02/04/2008
文件偏號: 3PD89BRM