How Do I Fix a Timeout Error Using the "TKTDS3K Getting Started" VI?

Primary Software: LabVIEW Development Systems>>Full Development System
Primary Software Version: 7.1
Primary Software Fixed Version: N/A
Secondary Software: Driver Software>>NI-VISA
Hardware: VXI/VME/MXI>>VXI Controllers

Problem: Why do I receive a timeout error (BFFF0015) when using the VI named "TKTDS3K Getting Started" to communicate with a TEK TDS30xx scope through a VISA Connections, and how do I fix it?

Solution: LabVIEW will wait on the correct values for the probe settings on the instrument, timing out if they are not received. Try setting the probe (10x or 1x) to the correct value for your particular system.