Which DIO lines are Reserved for Communication on a TEDS SC-2350 Carrier?Hardware: SCC>>Carriers>>SC-2350
Problem: When using TEDS class II sensors which DIO lines are available for use and which ones are reserved? Solution: With the SC-2350, DIO lines 0, 1, 2, and 4 are reserved for TEDS communication. If the SC-2350 device is configured in MAX, these 4 lines will not be available for use. All other DIO lines can be used for communication. In the case of the SC-2345, there are no reserved lines. Related Links: Product Reference: SC-2345/2350 Carrier User Manual Attachments:
Report Date: 09/27/2004 Last Updated: 10/27/2006 Document ID: 3DQDJH7U |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||