Easy Way to Test if My DAQCard 700 is Working Properly

Hardware: Legacy DAQ Devices>>Multifunction DAQ>>DAQCard-700

Problem: I have a DAQCard 700, and I am not sure if it is working properly. What is an easy way to test it?

Solution: DAQCard 700 requires NI-DAQ 6.9.3 drivers in order to function. If that is installed, open up Measurement and Automation Explorer and expand "Devices and Interfaces". The DAQCard 700 should be located there. Connect a terminal block to the DAQCard 700. The CB-50 is recommended. Take an analog input and connect it to AIGround. Connecting pin 1 and pin 3 will do on the CB-50. Open up test panel in Measurement and Automation Explorer and test channel 0. It should be reading 0 volts. If otherwise, then the card could be defective.

Related Links:

Attachments:





Report Date: 12/30/2004
Last Updated: 01/04/2005
Document ID: 3GTE0AV1