Triggering An Acquisition Session of a HSDIO Device Hardware: Digital I/O (DIO)>>High-Speed>>PCI-6552
Problem: I have configured my HSDIO acquisition session to have a digital start trigger. However, I am receiving a timeout error because it is not seeing the trigger. Why? Solution: A HSDIO acquisition session trigger must meet the setup and hold time requirements listed in the Acquisition Signal Characteristics section of the device specifications. Below is an image of two triggers. The first does not meet the setup and hold time requirements and would not start the acquisition. The second trigger meets the minimum setup and hold time requirements and would trigger the acquisition. The acquisition of data will begin on the same clock cycle in which the asserted start trigger meets the minimum setup and hold time requirements. Related Links: NI Digital Waveform Generator/Analyzer Help Attachments:
Report Date: 12/21/2006 Last Updated: 01/19/2007 Document ID: 44KF4H17 |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||