Error 1074135039 With Calibration Executive and PXI-4060Hardware: PXI/CompactPCI>>Controllers
Problem: I am using Calibration Executive 3.1 to calibrate my PXI-4060. My calibrator is the Fluke 5700A. When the procedure reaches the "AC Volts Verification" step, it gives error 1074135039. At this time, the 5700A displays a value of 237.5 V DC on its screen. What is the problem? Solution: This is a known issue fixed in Calibration Executive 3.1.2. You can download the latest version from the following site: Calibration Executive 3.2 As mentioned in the Readme file, this version modified the AC V test value from 237.5 V to 219 V so the procedure will work with all Fluke 57XX calibrators. Related Links: KnowledgeBase 3WCM871O: Error 1074135039 Occurred Using Calibration Executive Attachments: