Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
This Document is not yet Rated  Rate this Document

Error 1074135039 With Calibration Executive and PXI-4060

Hardware: PXI/CompactPCI>>Controllers

Problem:
I am using Calibration Executive 3.1 to calibrate my PXI-4060. My calibrator is the Fluke 5700A. When the procedure reaches the "AC Volts Verification" step, it gives error 1074135039. At this time, the 5700A displays a value of 237.5 V DC on its screen. What is the problem?

Solution:
This is a known issue fixed in Calibration Executive 3.1.2. You can download the latest version from the following site:

Calibration Executive 3.2

As mentioned in the Readme file, this version modified the AC V test value from 237.5 V to 219 V so the procedure will work with all Fluke 57XX calibrators.

Related Links:
KnowledgeBase 3WCM871O: Error 1074135039 Occurred Using Calibration Executive


Attachments:





Report Date: 10/18/2006
Last Updated: 03/17/2009
Document ID: 42H8K3E1

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit