Error -10411 with 1200 Series DAQ Board

Primary Software: Driver Software>>NI-DAQ
Primary Software Version: 6.9.3
Primary Software Fixed Version: N/A
Secondary Software: N/A
Hardware: SCXI>>Modules>>SCXI-1200, Legacy DAQ Devices>>Multifunction DAQ>>PCI-1200, Legacy DAQ Devices>>Multifunction DAQ>>Lab-PC-1200AI, Legacy DAQ Devices>>Multifunction DAQ>>Lab-PC+, Legacy DAQ Devices>>Multifunction DAQ>>DAQCard-1200, Legacy DAQ Devices>>Multifunction DAQ>>Lab-PC-1200

Problem: I receive error -10411 (counterBusyerror) with a 1200 series DAQ board (Lab-PC+, Lab-PC 1200/AI, DAQcard 1200, PCI 1200, SCXI 1200).

Solution: Two different operations are trying to access counter 0 (B0, the user-accessible counter that can access the internal timebase) at the same time. The counter B0 is used as an extended timebase for timed analog input or output when sample interval exceeds 65.535 ms.

This error can occur if you try an analog output or counter operation while performing a slow analog input operation. To resolve the error, speed up the sampling rate, so that counter B0 is not reserved by the analog input operation.