How Can I Connect Quater Bridge to NI9944 or NI9945 with NI-9237 on CompactRIO?

Hardware: CompactRIO>>Analog Input Modules>>NI 9237

Problem:
I physically connect a quater bridge to NI9944/9945 following the step in KnowledgeBase 47SGM5CY: NI 9945 or NI9944 with an NI 9237 Connection Guide.
However, the reading still doesn't make sense on CompactRIO system.
Do I have to configure anything in order to read the signal correctly?

Solution:
NI9944/9945 has a half-bridge circuit inside.
Therefore, Half-Bridge Completion must be enabled in order to measure quater bridge with NI 9944 or NI 9945 on Compact RIO system correctly.
Following below instruction to enable Half-Bridge Completion:
1. Right click on NI 9237 under FPGA or Chassis target
2. Select Properties from pull-down menu
3. Tick Enable Half-Bridge Completion on the channel connected to NI 9944 or NI 9945.
Attached screenshot shows how to enable half-bridge completion.



Related Links:
KnowledgeBase 47SGM5CY: NI 9945 or NI9944 with an NI 9237 Connection Guide
Product Manuals: NI 9237 Operating Instructions and Specifications


Attachments:
Half-Bridge Completionhb.PNG




Report Date: 09/21/2008
Last Updated: 12/19/2008
Document ID: 4PK5M1DQ