Calibration Executive: Which DAQ Devices Can Be Used to Calibrate SCXI Modules? Primary Software: Calibration ExecutivePrimary Software Version: 2.2.1 Primary Software Fixed Version: N/A Secondary Software: N/A
Problem: I need to calibrate my SCXI modules, yet with some DAQ devices my module fails. However, with a different DAQ device, my module passes the final verification of Calibration Executive. What DAQ device can I use to calibrate my modules? Solution: Only XE-10 series DAQ products can be used to calibrate SCXI products, excluding the SCXI-1124. These DAQ devices include:
*The 6031E/6033E have 64 input channels which require additional cables. Thus, the 6031E and 6033E are not the preferred devices for SCXI calibration. NOTE: M-series devices can be calibrated with Calibration Executive, however, they cannot currently be used to calibrate SCXI. Related Links: Calibration Solutions from National Instruments Attachments: