Verify Board Calibration Without Adjusting Calibration Constants

Primary Software: Calibration Executive
Primary Software Version: 2.2.1
Primary Software Fixed Version: 2.2.1
Secondary Software: N/A

Problem: Is there a way to verify that my data acquisition device is properly calibrated without adjusting the calibration constants stored in the device's EEPROM?

Solution: Yes, you can verify that your device is properly calibrated without running a complete calibration procedure that adjusts the calibration constants. Verify calibration by performing the following steps:
  1. Create or open a calibration procedure
  2. Select Execute » Run Mode » Verify Only


Related Links: KnowledgeBase 2Q8D29B9: Disabling Calibration Verification Using Calibration Executive