当我将计算机从睡眠状态唤醒时,如果我尝试在NI MAX或我的开发环境中访问NI-DAQmx硬件,则会收到错误-50152发生了硬件故障。无法按照指定完成操作。为什么会出现此错误,如何解决错误?
当计算机启动或从低功耗状态(例如睡眠)启动时,NI-DAQmx硬件必须完成自配置才能使用。在正常引导期间,操作系统加载所需的时间允许大量时间使硬件完成其自配置。然而,休眠状态将计算机的状态保留在存储器中,这导致负载更快地发生,特别是在使用诸如SSD的较快硬盘驱动器的系统上。由于自动配置在操作系统加载之前未完成,因此超时并导致-50152错误。
为了解决此错误,建议您重新启动计算机。为了防止将来发生错误,建议您不要将计算机置于睡眠电源状态,并禁用任何电源选项。
|