Delay and Error When Using NI 4451/NI 4452 DSA Board with VirtualBench-Logger Primary Software: Other NI Software>>VirtualBenchPrimary Software Version: 2.51 Primary Software Fixed Version: N/A Secondary Software:
Problem: When using my NI 44xx Dynamic Signal Analyzer (DSA) board with VirtualBench-Logger, it takes a long time for data to start appearing on the graph and before it does an error message appears. Solution: The NI 44xx DSA boards are different from the regular data acquisition boards and are not designed for low speed acquisitions, which VirtualBench-Logger is suited for. It takes a little while longer for the driver to configure the board to run a low speed (1 second time interval or greater, same as 1 scan per second or less). The error message is displayed because of this. The logging will take place but for slow logging it is better to use a different board if you are planning on using this software. Related Links: Attachments:
Report Date: 12/08/1999 Last Updated: 02/12/2003 Document ID: 1S7DGFTS |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||