Offset Seen Using Open TC DetectionHardware: PXI/CompactPCI>>Controllers
Problem: When I use open TC detection with TC-2095 and SCXI 1100, I see a offset of 2.5 VDC. How can I eliminate this? Solution: Open-thermocouple detection circuitry can cause two types of measurement errors:
With 10 MΩ pull-up and bias resistors, a common-mode voltage of 2.5 VDC develops if the thermocouple is floating. At a gain of 100, the common-mode rejection of the SCXI-1102/B/C module is sufficiently high that the resulting offset error is negligible. If the application demands extremely high accuracy, you can eliminate this offset error by:
The open-thermocouple detection circuitry results in a small current leakage into the thermocouple.
Product Manuals: BNC/TC-2095 Rack-Mount Adapter Installation Guide and Specifications Related Links: KnowledgeBase 2Y57ST7Y: Open-Thermocouple Detection BNC/TC-2095 Rack-Mount Adapter Installation Guide and Specifications Attachments:
Report Date: 01/12/2009 Last Updated: 06/03/2009 Document ID: 4TB5UUHV |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||