Is It Possible to Modify the Steps of Calibration Executive Procedures?Primary Software: Calibration ExecutivePrimary Software Version: 3.3 Primary Software Fixed Version: N/A Secondary Software: TestStand
Problem: I only want to calibrate certain parts of my hardware. Is there a way skip certain steps? Solution: Some users only want to run certain tests for their hardware. For example, if you only use the analog inputs on my E-series board, then you might not want to take the time to calibrate the digital or analog outputs. The calibration procedures are TestStand sequence files that step through the multiple calibration steps for each product. Since Calibration Executive ships with the TestStand RTE and not a full sequence editor, the procedures (TestStand Sequence files) cannot be modified with the software included. With a full version of TestStand, you could open the sequences and skip or change any part of the test they need. This would allow you to customize the steps that are performed during calibration. If a user opens the sequence file in TestStand and highlights the selected step, he/she has the option of setting the run mode. One of these options is Skip<. This will leave the step in the test in case he/she wants to use that functionality later, but it will skip over the individual step until it is set to run. Related Links: Products & Services: NI TestStand Products & Services: NI Calibration Executive Product Manuals: Calibration Executive Help Attachments:
Report Date: 12/05/2002 Last Updated: 02/17/2009 Document ID: 2S4GRBJY |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||