GPIB-RS232 or GPIB-RS485 Converter Communication in C ModeHardware: GPIB>>External Controllers/Convertors>>GPIB-RS232, GPIB>>External Controllers/Convertors>>GPIB-RS485
Problem: Why can't I communicate with my GPIB device after I set up my GPIB-RS232 or GPIB-RS485 converter in C Mode? Solution: In the NI GPIB-Serial Converter Wizard, you must select the primary address of your instrument (not the converter) as the GPIB Primary Address as shown in the screenshot below.
Note: C Mode can only be used if you have ONE GPIB instrument connected to the bus. Related Links: Product Support: GPIB-RS232 Product Support: GPIB-RS485 KnowledgeBase 4TJJ9L1N: Termination Modes for GPIB-RS232/GPIB-RS485 in C Mode Attachments:
Report Date: 01/20/2009 Last Updated: 08/14/2009 Document ID: 4TJIL41N |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||