Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
1 ratings:
 3 out of 5     Rate this Document

[c]FP-TC-120 Electromagnetic Immunity

Hardware: FieldPoint>>Analog I/O Modules>>FP-TC-120, Compact FieldPoint>>Analog I/O Modules>>cFP-TC-120

Problem: When the [c]FP-TC-120 is exposed to electromagnetic fields sensors may have voltages induced by such fields thus causing erroneous data.

Solution: To reduce the effect of electromagnetic fields use shielded thermocouple wires and install the
measurement system far from the source of the unwanted interference. Installing the measurement system in an EMC rated enclosure will also help protect the system from the unwanted interference. Proper filtering such as averaging can also help provide more accurate measurements.

Related Links:

Attachments:





Report Date: 04/04/2005
Last Updated: 04/19/2005
Document ID: 3K3DGKVM

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5
Please Contact NI for all product and support inquiries.submit