Error -88716 When Using A NI-DAQmx Device in LabVIEW Real-time With NI-DAQmx 8.7.1

Primary Software: Driver Software>>NI-DAQmx
Primary Software Version: 8.7.1
Primary Software Fixed Version: 8.7.1
Secondary Software: N/A

Problem:
When I try to self-test or use a DAQ device installed in a LabVIEW Real-time system with NI-DAQmx 8.7.1 I get the following error message:

Error -88716 occurred at Test Panel

Possible Reason(s):

The requested operation failed because the versions of a driver installed on the local and the remote system are different.

Update the driver so the versions on the two systems are the same.


Solution:
NI-DAQmx 8.7.1f2 has a known issue that causes this error.  NI-DAQmx 8.7.1f3 resolves this issue.  If you are getting this error confirm that you have NI-DAQmx 8.7.1f2 installed and then download and install NI-DAQmx 8.7.1f3.  After downloading and installing NI-DAQmx 8.7.1f3 update NI-DAQmx on the LabVIEW Real-time target.

To determine which version of NI-DAQmx is currently installed on your host system open Measurement and Automation and expand the Software section under My System.  Select NI-DAQmx to read the version.
Determining NI-DAQmx Version


Related Links: