This content is not available in your preferred language.

The content is shown in another available language. Your browser may include features that can help translate the text.

Count Digital Edges Using the USB-6008/9 in LabVIEW

Updated Apr 24, 2023

Reported In

Hardware

  • USB-6008
  • USB-6009

Issue Details

I have a USB-6008 or USB-6009 and I want to be able to count digital edges or events in LabVIEW. I have tried using the Counter Input example programs in LabVIEW but no matter how I configure the channel settings I get error -200077 saying the "requested value is not a supported value for this property."

Solution

You can configure PFI 0 as a source for counting digital edges. In this mode, falling edge events are counted. DAQmx Create Channel function has a terminal for edges that defaults to rising if unwired. When using the USB-6008/9 you must create a constant or control that sets the channel to count falling edges. 

Additional Information

In the LabVIEW Counter - Count Edges.vi example program, the active edge being counted is changed using a property node. Even if you set the value of this property to be falling, LabVIEW still gives an error when using the USB-6008/9. This is because the DAQmx Create Channel function is called first and it defaults to the rising edge when nothing is wired. To solve this issue you must also wire a constant or control into the edge terminal of the function and set it to falling. 

Was this information helpful?

Yes

No