Why Do I Get Error –1074118135 When Performing Calibration on NI-4070/4072/4072B?

Hardware: Modular Instruments>>Digital Multimeters (DMM)>>PXI-4070, Modular Instruments>>Digital Multimeters (DMM)>>PXI-4072, Modular Instruments>>Digital Multimeters (DMM)>>PCI-4070

Problem:
Why am I getting error –1074118135 when calibrating a NI 4070/4072/4072B using Calibration Executive?

Solution:
Error -1074118135 indicates Invalid leakage resistance. This error can occur during adjustment when the NI-DMM circuitry is too hot. Make sure to follow the instructions in the Test Conditions section of Calibration Executive Help to maintain a proper test environment. From Calibration Executive Help, Browse to Calibration Executive Help » Device Calibration Procedures » Modular Instruments » Digital Multimeters » NI-4070/4072 Calibration Procedure.

Related Links:
NI PXI-4070 PXI DMM
Calibration Executive Help
NI 4070/4072 61/2 Digit FlexDMM Calibration Procedure

 



Attachments:





Report Date: 04/10/2006
Last Updated: 05/06/2008
Document ID: 3W9BFRMO