Does the SCC-A10 Require 4 Analog Input Channels?Hardware: SCC>>Modules>>SCC-A10
Problem: I see that there are 2 differential channels on the SCC-A10, does this mean I have to use 4 analog input channels on my DAQ board? Solution: The SCC-A10 has two Differential analog input channels, but the Differentails are measurred inside the module. Therefore, each module only needs need 2 analog inputs and not 4. Also, note that due to this fact, the signal will be recognized as a NRSE signal even though it has been wired as a differential to the SCC-A10. Please see diagram bellow to see how the module functions internally.
Related Links: Product and Service: SCC-A10 Attachments:
Report Date: 11/03/2008 Last Updated: 11/04/2008 Document ID: 4R2K4ESL |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||