SC-2043-SG Output Buffers Oscillate When Used with 5 m and 10 m Cables

Hardware: Signal Conditioning Other>>SC Analog Accessories

Problem: When the SC-2043-SG is used with the 5 m or 10 m shielded cables, the output buffers are likely to oscillate due to instability caused by excessive cable capacitance.

Solution: Use a shorter cable or contact National Instruments technical support (see the Request Support link, below) and ask about an ECO update for the SC-2043-SG. Please refer to this KnowledgeBase entry (13IG8H41) in your communication.

Related Links: Request Support