Problem: What is the maximum update rate of VXI-AO-48XDC?

Solution: The VXI-AO-48XDC has one 18-bit DAC which is time-division multiplexed to create the 48 voltage and 48 current output channels. Thus, a channel can take a duration a maximum of 5 ms to change value after it has been updated by software; the update rate is about 200 points/second.