为什么NI 5752的输入被反转了?



硬件: Modular Instruments>>High-Speed Digitizers (Scopes)>>NI 5752

问题: 当我用NI 5752获取数据时,数值被反转了。这是预期之内的吗?我要怎么解决?

解答: NI 5752是基于TI AFE5801的,它是一个反相差分放大器。CLIP会输出一个反转信号给LabVIEW。为了解决这个问题,可在CLIP输出后面加一个二进制反转。

相关链接: External Link: AFE5801 Datasheet