Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
3 ratings:
 4.66 out of 5     Rate this Document

Offset Nulling with the SCC-SG24

Hardware: SCC>>Modules>>SCC-SG24

Problem:
I am using the SCC-SG24 module for strain measurements. When I use the potentiometers to null the offset of my strain gage, the signal remains the same. The offset is still present regardless of how I adjust the nulling potentiometer. What am I doing wrong?

Solution:
By default, the offset nulling resistors are not installed on the SCC-SG24. These resistors ship with the module, but you will need to install them to get the offset nulling potentiometers to work. Please see the "Replacing Resistors" section of the SCC-SG Series User manual linked below.

The reason why these resistors are not installed at factory is because the SCC-SG24 module is usually use for Load Cell applications. Generally, there are a large number of load cells that are not compatible with offset nulling resistors because they would introduce an unwanted offset in the measurements.

Related Links:
Product Reference: SCC-SG Series Strain-Gauge Modules User Guide
Developer Zone Tutorial: Measuring Strain with Strain Gauges

Attachments:





Report Date: 09/04/2004
Last Updated: 02/22/2008
Document ID: 3D3E39W1

Your Feedback! poor Poor  |  Excellent excellent   Yes No
 Document Quality? 
 Answered Your Question? 
  1 2 3 4 5